:MTESt:AMASk:CREate

Command Syntax
:MTESt:AMASk:CREate
The :MTESt:AMASk:CREate command automatically constructs a mask
around the current selected channel, using the tolerance parameters
defined by the :MTESt:AMASk:XDELta, :MTESt:AMASk:YDELta, and
:MTESt:AMASk:UNITs commands. The mask only encompasses the portion
of the waveform visible on the display, so you must ensure that the
waveform is acquired and displayed consistently to obtain repeatable
results.
The :MTESt:SOURce command selects the channel and should be set before
using this command.
See Also
Example Code
Agilent InfiniiVision 2000 X-Series Oscilloscopes Programmer's Guide
(see
page
790)
"Introduction to :MTESt Commands"
":MTESt:AMASk:XDELta"
":MTESt:AMASk:YDELta"
":MTESt:AMASk:UNITs"
":MTESt:AMASk:SOURce"
":MTESt:SOURce"
on page 409
"Example Code"
on page 379
on page 379
on page 386
on page 387
on page 385
on page 384