When Ordered as an Upgrade from a Smaller Solid State Disk Drive

1. With the original solid state disk drive still installed in the instrument, determine what version
of the instrument software is currently installed on it. For instructions on how to verify the
currently installed software version see the
instructions. Record the currently installed software version below.
Instrument S/W Revision: ______________________
2. Backup the instrument factory calibration data to a USB storage device. See the
or Restore using Alignment Data Wizard (for processor type PC8 or PC9)
document for instructions.
3. Determine FPGA Configuration (if possible). Press System, on left side pane select Power On.
Look for FPGA Configuration selection to appear. If FPGA Configuration does not appear
continue on. If FPGA Configuration selection appears, check the box below corresponding to
the currently loaded FPGA. See
[ ] Enhanced Sweep Speed
[ ] Time Domain Scan
Figure 1
FPGA Configuration
4. Replace the original instrument disk drive with the solid state disk drive supplied with this kit.
See the
Disk Drive Removal and Installation
5. Turn the instrument on with the new solid state disk drive installed and see the
Instrument Operating System
6. Ignore any hardware related error messages that might appear at this time, as they may be
related to software version incompatibilities that will be resolved in the following steps of this
procedure.
Installation Note N9032-90007
Software Updates
Figure
1.
section in this document for instructions.
section in this document for instructions.
Solid State Disk Drive
section in this document for
Data Backup
section in this
Initializing the
7