Correction Constants; Storing Correction Constants - Keysight Technologies N5247B Service Manual

2-port and 4-port pna-x microwave network analyzer
Hide thumbs Also See for N5247B:
Table of Contents

Advertisement

General Purpose Maintenance Procedures

Correction Constants

Correction Constants
The analyzer stores many correction constants in non-volatile EEPROM
memory. These constants enable the analyzer to produce accurate, leveled
source signals and receive clean test signals.

Storing Correction Constants

After performing any adjustment listed on
page 3-54
in this manual, store the
correction constants to a backup file on the analyzer solid state drive by
performing these steps:
— Navigate to the EEPROM Backup Utility, located at:
C:\Program Files (x86)\Keysight\Network Analyzer\Service\eebackup.exe
— Run the program.
— Click Backup EEPROM.
— Click Backup TSMB Mem.
— Click Exit when the program has finished.
8-16
Keysight N5247B Service Guide