To Connect The Gndu Output - Keysight Technologies E5260 Series User Manual

Hide thumbs Also See for E5260 Series:
Table of Contents

Advertisement

Kelvin connections
Use a low-noise coaxial cable (Keysight part
number 8121-1189) from the connector to the
prober, socket, or DUT as shown.
To cancel the effects of cable resistance, connect
the Sense line and Force line as close as possible to
the terminal of the DUT.
Common
Force
Sense
Insulator
Plate
Triaxial connector
CAUTION

To Connect the GNDU Output

The following figures show connection examples for the GNDU output.
to
Coaxial cable
Use Keysight 16493L GNDU cable to connect the GNDU to your connector plate or
test fixture. Do not use the triaxial cable. The GNDU is rated for up to 4 A
(E5260A/E5270B) or 2.2 A (E5262A/E5263A), while the maximum current rating
of the triaxial cable is 1 A.
Keysight E5260/E5270 User's Guide, Edition 6
non-Kelvin connections
Short the Sense and Force on the connector as
shown below. Measurement data will include the
residual resistance of the connection wire.
Use AWG 22 single-strand insulated wire
(Keysight part number 8150-2639) from the
connector plate to the prober, socket, or DUT.
For a quick connection where measurement
accuracy is not critical, connect only Force to the
DUT, without shorting the Sense and Force. With
this connection, the measurement data will include
residual resistance from the connection cable
between the GNDU and the connector plate.
Common
Force
DUT
Plate
Triaxial connector
Mounting Connectors
Sense
Insulator
Wire
Installation
to
DUT
3-29

Advertisement

Table of Contents
loading

This manual is also suitable for:

E5270 seriesE5260aE5262aE5263aE5270b

Table of Contents