Flow Tracker Series 1000/2000
Flow Tracker 1000/2000 Flow Measurement Specifications
Table 6
Flow measurement specifications
Flow range for measurement
Accuracy
(over 5–500 mL/min)
Sensor type
Calibration
Calibrated gases
Modes
Display
Power
Dimensions
Certification
Operating pressure
Output
Weight
Inlet and outlet
Accessories
* Median reading using test procedure. Test procedure can be found at: www.agilent.com/chem/supplies under the gas
management section
34
Table 6
provides flow measurement specifications for the Flow
Tracker 1000 and Flow Tracker 2000 flowmeters.
0–500 mL/min
Flow:
Temp:
Press:
Flow:
Temp:
Press:
NIST traceable multi-point calibration certificate included
Nitrogen, hydrogen, helium, air, carbon dioxide, methane,
95% argon / 5% methane blend
Flow, gas chromatography, leak detection (Flow Tracker 2000 only)
Dynamically labeled multi-function 7 line LCD with 99 characters
Six AA batteries (included) or optional universal AC/DC Adapter
(Part number 5183-4781)
8.3-inch (h) x 3.75-inch (w) x 1.9-inch (d)
CE marked
100 PSI (690 kPa) maximum
RS-232 output for all variables for selected mode, no special software required
Approx. 2.4 lbs (1.1 kg)
1/8-inch NPT (female) threads
RS-232 cable, stand, batteries, 1/8-inch and 1/4-inch barbed fittings
± 2% of reading, ± 0.2 mL/min
*
(± 5% from 0–5mL/min)
± 0.5 °C
± 0.5% of reading
Solid-state piezoresistive differential pressure sensor
with measurement in the laminar flow region. No
heated elements
Integrated circuit absolute temperature sensor. No
RTD's
Solid state absolute pressure sensor
Reference Manual