SpO2 Measurement (optional)

Sensor selection

Only use Nellcor sensors.
Note the Instructions for Use of the sensors –
incorrect positioning or use can cause tissue damage.
Select the sensor according to the following criteria:
– Patient weight
– Patient mobility
– Possible application site
– Patient perfusion
– Duration of use
The following table listing the specific sensor types
available and their characteristics is intended as an aid
for selection.
Sensor type
OXISENSOR
I-20
Age group
Infants
Patient
1 to 20 kg
weight
Duration
Short and long-term Short and long-term Short-term
of use
monitoring
Patient
Limited
mobility
mobility
Preferred
Toe
measuring site
1)
Sterility
Sterile packaging
OXISENSOR ™ I-20, OXISENSOR ™ D-20,
DURASENSOR ™ DS-100 A, OXISENSOR ™ D-25 and
OXISENSOR ™ R-15 are registered trademarks.
–––––––––––––
1) in unopened, undamaged packaging
76
OXISENSOR
DURASENSOR OXISENSOR
D-20
DS-100 A
Children
Adults
10 to 50 kg
>40 kg
monitoring
monitoring
Limited
Immobile
mobility
patients only
Finger
Finger
Sterile packaging
____
OXISENSOR
D-25
R-15
Adults
Adults
>30 kg
>50 kg
Short and long-term Short and long-term
zeitüberwachung
monitoring
Limited
Immobile
mobility
patients only
Finger
Nose
Sterile packaging
Sterile packaging