Continuous Recordings

Continuous recordings provide an up-to-the-second printout of
patient data. This mode is useful when more than 20 seconds of
data must be printed during critical, short-term applications.
NOTES:
!
!
STEPS: Starting a Continuous Recording
1. Press the
beeps indicate when the recording starts.
2. To cancel the recording, press the
the
Continuous recordings cannot be stored. If you attempt to store a
continuous recording, the monitor will store it as a timed record-
ing with 20 seconds of patient data.
NOTES:
!
!
Page 7-8
You can request a continuous recording during an alarm, timed,
trend, and diagnostic log recording. To do so, first stop the
recording in progress with the Record fixed key on the moni-
tor, then press and hold the same key to start the continuous
recording.
Continuous OCRG recordings are not supported.
fixed key for at least two seconds. Two
Record
key of the recorder while printing is in progress.
Stop
If the user requests a continuous recording at the central sta-
tion while a timed recording is in progress, the timed recording
is cancelled. A continuous recording can only be printed out if
two recorders are available. If only one recorder is available,
the continuous recording request is converted to a timed
recording, then stored and printed when the recorder becomes
available.
An uninterrupted continuous recording on a laser printer con-
sists of up to 120 pages.