Continuous Positive Airway Pressure Cpap - In Pressure Support Mode (Optional) - Dräger Primus Infinity Empowered Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Primus Infinity Empowered:
Table of Contents

Advertisement

Press. Support
Apnea ventilation can additionally be set by means
of the minimum frequency Freq
lator is automatically triggered by Freq
no spontaneous breathing activity by the patient.
This is not a mandatory ventilation stroke by the
ventilator; the patient can end the stroke triggered
by the ventilator at any time by breathing spontane-
ously. This stroke is not identified by a trigger indi-
cator.
Apnea ventilation can also be deactivated again by
means of Freq
(OFF position).
MIN
Continuous Positive Airway Pressure
CPAP – in Pressure Support Mode
(optional)
Spontaneous breathing at an increased pressure
level to increase the functional residual capacity.
Continuous Positive Airway Pressure is activated in
Press. Support when the value for pressure sup-
is set to ≤2 [hPa (cmH
port
P
PS
At the same time, the minimum frequency Freq
is set to OFF and the rise time T
Apnea ventilation with minimum frequency is not
appicable in the CPAP mode.
Instructions for use Primus Infinity Empowered SW 4.5n
B A
C
(C). The venti-
MIN
if there is
MIN
O)].
2
is set to 0.0.
SLOPE
Setting ranges and factory settings
Ventilation
parameters
Minimum fre-
2)
quency
Freq
MIN
[1/min]
PEEP
[hPa (cmH
O)]
2
Trigger sensitivity
Trigger
[L/min]
Pressure Support
P
PS
[hPa (cmH
O)]
2
Rise time
T
SLOPE
[sec.]
age
[years]
weight
[kg]
[lbs]
1) The default values can be set specifically for the hospital
MIN
concerned, see page 166.
2) The inspiratory time is limited by adjustment of Freq
to yield a maximum ratio of 1:1 for I:E, thus ensuring an
adequate expiratory time.
Operation
Setting
Factory
ranges
setting
OFF, 3 to 20
OFF
(CPAP)
0 to 20
0.3 to 15
3.0
>2 to 50
0 to 2
(CPAP)
0.0 to 2.0
0.0
0.0
(CPAP)
<1 to 120
40
1 to 120
– –
1 to 240
1)
3
0
5
MIN
115

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sw 4.5n

Table of Contents