NanoSuite Reference

Required Inputs

NOT E
Agilent Nano Indenter G200 User's Guide
There are some parameters in the NanoSuite software that are used only
with the CSM option.
When using the CSM option, the Required Inputs should display the
following five parameters as shown in
Figure A-4Required Inputs for CSM
When you conduct CSM testing, set the depth limit larger than you
want. As the indenter penetrates the sample, NanoSuite continuously
gathers data. If the indenter penetrates beyond a useable limit, the data
that NanoSuite collects before that point are still valid.
Depth Limit
Maximum penetration of the indenter into the test surface in nm.
Strain Rate Target
Loading is controlled such that the loading rate divided by the load on
the test surface is held constant at this value (1/s). This means that when
the load on the sample is small, the loading rate is small, and as the load
increases, the loading rate increases.
The default is 0.05/s. For example, if the absolute load is 1 mN, then the
loading rate is 0.05 mN/s at the default value.
Continuous Stiffness Measurement A
Figure A-4
and described below.
A-5