Second Batch Test

Agilent Nano Indenter G200 User's Guide
The Statistics panel at the bottom of the screen displays results from
each test. The statistics should be similar to those shown in
Note the specific results to be displayed are determined by instructions
in the method. Take a few minutes to review the results for each test.
Figure 2-66Statistics panel
This part of the tutorial uses a default method and batch to perform a
series of tests on fused silica. The default method prescribes a test that
returns the Young's modulus and hardness at a series of discrete loads.
The default batch prescribes a 2 x 5 array of these tests on a single
sample. For more information on the fundamental analysis used to
derive Young's modulus and hardness from indentation data, consult
Chapter
7, "Theory".
Click the Test tab.
1
Click the icon for the Batch Mode. The default batch for the
2
instrument will automatically be loaded in the right panel.
Begin by clicking the Run Batch button.
3
In the resulting Review Sample Names dialog box, type in fused
4
silica 2 for the sample name. Click the OK button to continue.
In the Review the first test location of Sample dialog box, move the
5
sample to the correct location under the microscope. Right-click
anywhere in the video screen in the dialog box and select Nano
HandSet from the resulting pop-up menu.
Using the schematic map, click on the location where the fused silica
6
sits in the sample tray (click on one of the circles).
Right-click on the marked circle and select Move to Target from the
7
resulting pop-up menu.
Right-click anywhere on the schematic map and select Nano Video
8
Handset from the resulting pop-up menu.