NanoSuite 4
Figure 4-50Nano Handset with icons for Target location and Current
Position
A green icon represents the completion of a move to the new location,
as shown in
Figure
4-51.
Figure 4-51Completed Move to Target icon
Load Sample Tray
This command moves the sample-tray retainer all the way forward to
enable you to remove or install a sample tray. With the tray in this
position, the risk of damaging the indentation system is minimized.
Initialize
This command is used to reset the motion system to its initial settings.
The system will accomplish this by moving the X/Y table until the
reference limit switches are found. This command is helpful when the
actual position of the sample tray with respect to the microscope does
not correlate to the position indicated on the schematic.