Figure 1-3
Before proceeding to the reported self-test error code, check the voltages in Table1-18. If any of
the voltages are out of specification troubleshoot the supply problem first.
Table 1-18
Supply
Voltage
(Vdc)
300 1 GHz Detector
1. Check and remove any connection to the 10 MHz external reference on the rear panel.
2. If the signal generator continues to fail self-test, replace the A7 Reference.
301 Tuning Voltage
1. Replace the A7 Reference.
1-26
Connector
Pins
+32
P102-7
+15
P102-6, 21
-15
P102-2, 17
+5.2
P102-14, 29
+9
P102-5, 20
Minimum
Maximum
Value
Value
(Vdc)
(Vdc)
+31.04
+32.96
+14.55
+15.45
-14.55
-15.45
+5.04
+5.37
+8.82
+9.18
Origin
Main Supply
Main Supply
Main Supply
Main Supply
YIG Driver