Comparing Samples In Analyst - Agilent Technologies Nano Indenter G200 User Manual

Table of Contents

Advertisement

Comparing Samples in Analyst

Agilent Nano Indenter G200 User's Guide
NanoSuite does not compare multiple samples. This procedure must be
carried out in Analyst.
Launch NanoSuite.
1
Open the first sample that you want to include in your comparison.
2
Select Output Sample to Excel from the Export menu.
3
Repeat
step 1
through
4
include in your comparison.
Launch Analyst.
5
Select Calculations from the Options menu.
6
If the data were collected using the basic instrument, then select
a
the Create Sample Summary Sheet radio button in the
Calculations dialog box.
If the data were collected with the CSM option, then select the
b
Create Sample Average Sheet radio button as shown in
Figure
6-98.
Refer to the
Options
menu in
about the Calculations dialog box.
Figure 6-98Calculations dialog box
Commonly Used Procedures 6
step 3
for each sample that you want to
Chapter
5, "Analyst" for more information
6-73