Agilent G2545A Hybridization Oven
Calibration Procedure
Step 1. Set up and heat the oven
Step 2. Start the digital recording thermometer
Step 3. Retrieve the temperature data from the digital recording
thermometer
Step 4. Calibrate the oven
The Agilent G2545A Hybridization Oven maintains temperature calibration
for at least 3 months under normal operating conditions when properly
installed within the laboratory.
See the Agilent G2545A Hybridization Oven Installation, Operation, and
Maintenance Guide (p/n G2545-9001) for information on site
requirements and oven placement.
Calibrate the Agilent G2545A Hybridization Oven:
• After first installation in a working environment
• After each 3 months of use
• When systematic high microarray background noise may indicate a drift
in oven temperature
Do the calibration steps in this guide to maintain optimal performance for
your Agilent G2545A Hybridization Oven.
To view a demonstration of the hybridization oven
calibration procedure, go to
http://www.agilent.com/genomics/protocolvideos, or scan this
QR code with your mobile device.