Agilent Nano Indenter G200 User's Guide
The reduced modulus of fused silica is known because the properties are
known:
For diamond, E = 1140 GPa and n = 0.07. For fused silica, E = 72 GPa
and n = 0.18, yielding:
1 0.18
E
=
-------------------------- -
r
The following parameters are needed for semi-static methods for
determination of indenter area function.
Contact Stiffness is defined as:
Contact Depth is defined as:
Contact Area is defined as:
Each indentation provides one ordered pair of (h
Figure 7-12
on page 7-20.
2
2
1 0.07
+
-------------------------- -
72
1140
dP
S
=
------ -
dh
h
max
0.75P
max
h
=
h
---------------------- -
c
max
S
2
S
-- -
-----
A
=
4
E
r
Theory 7
=
69.87GPa
(23)
(24)
(25)
(26)
,A) as shown in
c
7-19