Agilent Nano Indenter G200 User's Guide
Figure B-23Open a DCM method
Install the sample tray, with the fused silica standard in the center
3
hole, into the instrument according to
the Indenter"
on page 2-18.
Set up a batch of samples for testing:
4
Select Batch Mode from the Mode menu.
a
Follow the instructions for
b
Chapter
2, "Quick Start Guide" using the fused silica in the center
of the sample tray. Follow the procedure exactly as written. In the
Required Inputs section, keep the default values and click Next
Step.
Click the Review tab to reveal the test results.
c
Right-click in the Legend pane and select Select All Tests.
d
Right-click on the graph and select Y-Axis Channel >
e
Modulus (GPa), as shown in
Dynamic Contact Module II B
"Install the Sample Tray into
"First Batch Test"
on page 2-27 in
Figure B-24
on page B-17.
B-16