Agilent Technologies Nano Indenter G200 User Manual page 147

Table of Contents

Advertisement

NanoSuite 4
Figure 4-59LED Settings dialog box
Microscope to Indenter Calibration
This command calibrates the distance between the objective and the
indenter tip, enabling you to precisely place indents. This calibration
should be done whenever you change the objective or tip.
Before selecting this command, the locking pins must be removed from
CAUTION
the indenter shaft.
This calibration should be performed using the single-crystal aluminum
sample provided with the machine. Before initiating this command, the
microscope must be focused on a clean spot of a soft-metal sample so
that the impression from the indenter will be easy to find.
Agilent Nano Indenter G200 User's Guide
4-55

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents