Print Preview of Tip Configuration
Change Indenter Tip
Agilent Nano Indenter G200 User's Guide
This command displays a Notepad text file listing all pertinent tip
information for viewing before a print is made. Choose Print from the
Notepad File menu to print this information.
This command initiates the procedure for changing the indenter tip.
After selecting this command, three dialog boxes will appear, each
requiring you to click the OK button:
The first step parks the head and moves the stages into a position that
1
allows easy access to the tip.
The second step asks you to ensure that the locking pins are in place
2
in the indenter shaft before performing a tip change.
The third step asks you to change the tip.
3
For more information, refer to
"Commonly Used Procedures".
Changing the Indenter
in
NanoSuite 4
Chapter
6,
4-33