Agilent Technologies Nano Indenter G200 User Manual page 333

Table of Contents

Advertisement

Agilent Nano Indenter G200 User's Guide
Once the method has been opened, continue with the procedure:
Install the sample tray, with the fused silica standard in the center
4
hole, into the instrument according to
the Indenter"
on page 2-39.
Set up a batch of samples for CSM testing:
5
Select Batch Mode from the Mode menu.
a
Follow the instructions for
b
Chapter
2, "Quick Start Guide" using the fused silica in the center
of the sample tray. Follow the procedure exactly as written. In the
Required Inputs section, keep the default values and click Next
Step.
Click the Review tab to reveal the test results.
c
Right-click in the Legend pane and select Select All Tests.
d
Right-click on the graph and select Y-Axis Channel >
e
Modulus (GPa), as shown in
Figure A-3Set Y-axis channel to Modulus
Zoom in and verify that the modulus is constant.
f
Right-click on the graph and select Y-Axis Channel >
g
Hardness (GPa).
Zoom in on the hardness.
h
Continuous Stiffness Measurement A
"Install the Sample Tray into
"First Batch Test"
on page 2-27 in
Figure
A-3.
A-4

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents