Previously, you modified only how data are analyzed. In this exercise,
you will modify the test flow to change how the experiment is
performed. Beginning with the standard CSM method, you will create a
new method that uses a constant loading rate to get to a target force
quickly and then uses a constant strain rate to finish the experiment. You
will use 50 mN for the target force and 10 seconds for the ramp time.
You will also change the default oscillation amplitude to 10 nm.
Launch NanoSuite, logging in as a user with Creator priveleges.
1
Choose Method > Open Method.
2
Select XP > G-Series CSM Standard Hardness, Modulus, and
3
Tip Cal from the Filename drop-down menu.
Click OK.
4
Select Save Method As... from the Method menu.
5
In the Save Method File As dialog box, type in G-Series CSM
6
Quick Ramp Then Constant Strain Rate for the file name.
Select Modified Methods from the Save in drop-down menu.
7
Click Save.
8
Click the Test tab to reveal the Test page.
9
Click the Define tab to reveal the Define page.
10
Click the Configuration tab.
11
Click the Inputs folder.
12
In the Settings pane, right-click and select Add Input... from the
13
resulting pop-up menu.
In the Insert New Variable dialog box, enter the internal name
14
RampForceLimit (no spaces).
Click OK.
15