Figure 6-83Choose the calibration method
Click Calculate.
28
Select Tip Area Results tab.
29
Coefficient: Best fit coefficients for A=m
Chi-squared: Square of the difference between fit and the data,
summed over all of the data. This value is minimized by the
fitting process, but it is not very useful for evaluating the quality
of the fit
Iterations: Number of iterations for convergence
K(
): Frame stiffness
LF
E
: Modulus used for determination of area function
A
Evaluate area function by verifying weighted difference of contact
30
area is centered about zero for all displacements. Deviation should
Commonly Used Procedures 6
2
d
+ m
d + m
0
1
0.5
d
+ ...
2