Agilent Technologies Nano Indenter G200 User Manual page 397

Table of Contents

Advertisement

Agilent Nano Indenter G200 User's Guide
Figure D-41Advanced tab for Channels
In the Notation drop-down menu, choose to display values in either
fixed or scientific notation. The Effect of Increasing Extension
drop-down menu is not used for indentation testing.
The following checkboxes turn these features On (select) or Off (clear):
• Calculate during Test: Controls whether the channel is calculated in
real time while the test is running. If you choose not to include the
channel during the test, you can include it in a calculation after the
test.
• Recalculate Post-Test: Controls whether the channel is calculated
after the test. This feature works in conjunction with the previous
feature. If this checkbox is selected, then selecting Recalculate
Sample or Recalculate Test from the Tools menu will recalculate
the channel.
• Never Recalculate: Controls whether the variable can be calculated
more than one time. If this checkbox is selected, then an option for
saving the raw data appears below the Effect of Increasing Extension
drop-down menu. However, this option does not apply to indentation
testing.
• Hidden: Controls whether the channel is hidden from view in the
Test and Review pages.
• Locked: Controls whether the channel can be deleted. If this
checkbox is selected, then the channel can be deleted only by
someone with Creator privileges.
• Allow Formula Editor: Controls access to the formula editor for
users with lesser privileges.