Agilent Nano Indenter G200 User's Guide
Click and drag on the sample image in the video screen to move the
9
sample under the fixed microscope to find a clear location for testing
(one without specs or dust). Click the OK button.
Close the machine cabinet.
10
Again, while the batch is running, note the numerical readouts, the
real-time graph, and the text below the real time graph. It should take
about thirty minutes to complete the batch.
When the batch is complete, the real-time graph is replaced by a table,
and the residual indentation impressions should be visible on the video
display.
If you are using a DCM, residual impressions may not be visible.
Reviewing Data from the Second Batch
You have just completed the second batch in which you performed a
series of ten tests on fused silica. Now, review the results.
Click the Review tab.
1
In the Legend panel, right-click and select Select All Tests from the
2
resulting pop-up menu.
This causes the Review Graph to display results from all ten tests. The
Review Graph now displays the calculated Young's modulus at the
discrete peak displacements for each loading cycle.
Right-click on the Review Graph and select X-Axis Channel >
3
Load On Sample (mN) from the resulting pop-up menu, as shown
in
Figure 2-67
on page 2-49.
Now the Review Graph displays the sample information as a function of
load rather than displacement.
Quick Start Guide 2
2-48