CAUTION
Agilent Nano Indenter G200 User's Guide
Figure 2-30Microscope focus and brightness controls
Do not adjust the focus controls in NanoSuite from this point forward
until the samples in the four corners of the sample tray have been set to
their proper heights, described next. Doing so could damage the
indenter tip.
Position Sample Under the Microscope
In handset area in the left panel, right-click and select Nano
1
HandSet from the resulting pop-up menu.
Click on the top left circle in the schematic map, which corresponds
2
to the rear left sample in the sample tray.
Right-click anywhere on the schematic map and select Move to
3
Target from the resulting pop-up menu.
Visually confirm that the microscope is over the rear left sample disc
4
in the sample tray. The microscope does not have to be directly over
the sample at this point.
Adjust the sample position until the light from the microscope shines
5
directly on the sample.
Locate the height-adjustment thumbwheel for the rear left sample disk
in the sample tray.
Figure 2-31
which is not visible from the front of the instrument. To access the
thumbwheel, you will have to reach behind the sample tray.
Quick Start Guide 2
on page 2-25 shows the thumbwheel,