Theory

Dynamic Modeling of the Indentation System and Sample Contact

Agilent Nano Indenter G200 User's Guide
The success of CSM relies on your ability to accurately model the
dynamic response of the indentation equipment so as to allow the
isolation of the material response. Nano Indenter systems are
purposefully designed so that the dynamic response of the indentation
equipment is easily modeled.
A schematic of an indentation head from a Nano Indenter system is
shown in
Figure
A-5.
Figure A-5Schematic of Nano Indenter indentation head
All motion is restricted to one dimension so that the system is well
modeled by a simple harmonic oscillator.
Each component in
Figure A-5
oscillator models in
Figure A-6
oscillator on the right is used to model dynamic response of a Nano
Indenter head in contact with a test sample.
Continuous Stiffness Measurement A
is represented in the simple harmonic
on page A-9. The simple harmonic
A-8