Standard Reference Materials

Temperature Probe

Agilent Nano Indenter G200 User's Guide
Your G200 is provided with two standard reference materials: Corning
7980 (fused silica) and Borofloat. These materials include calibration
certificates, on the left in
Young's modulus and Poisson's ratio that were determined by
ultrasound. Each block is marked with a serial number, as shown on the
right in
Figure 3- 23
on page 3-19. The replacement calibration
certificate can be obtained from Agilent with this serial number. These
standard reference materials are used to indirectly verify the instrument
according to ISO-14577.
Figure3- 23Standard Reference Material with calibration certification
(left) and serial number marker (right)
The test methods that comply with ISO 14577 acquire data from the
temperature probe inside the G200 cabinet, as shown in
Instrument Reference 3
Figure 3- 23
on page 3-19, that provide the
Figure 3-
24.