Agilent Nano Indenter G200 User's Guide
Post-Test Analysis to Determine Instrument Frame Stiffness
and Modified Results:
If this test method is being used to generate the necessary data for
determining instrument frame stiffness in situ, then the reported
properties, especially Young's modulus, should not be given much
weight until AFTER the calibration is complete and the results are
re-calculated using the new frame stiffness.
To perform the calibration:
From the main menu, select Tip > Configure Tip as shown in
1
Figure
H-10.
Figure H-10Select Configure Tip from the Tip menu
If prompted to add the current tip to your setup, click Yes.
2
In the Configure Tip Parameters window, click Calibrate Frame as
3
shown in .
Express Test I
H-14