Agilent Nano Indenter G200 User's Guide
Understanding the Printed Report
Most of the NanoSuite methods supplied with the instrument
automatically produce a printed report. Both the format and the content
of the automated report are prescribed within the method.
If your report did not print, you can display the report on your screen by
selecting Print Preview > Sample from the File menu, as shown in
Figure
2-71.
Figure 2-71View report by selecting Print Preview > Sample from the
File menu
From the printer, retrieve the printed report and review the following:
Table of test results: This is a tabulation of important results for
each test. Exactly which results are displayed is prescribed by the
method. At the bottom of the table, you will see statistics
compiled for all the tests on the sample.
First graph: This is a plot of Modulus vs. Displacement Into
Surface for all tests performed on the sample.
Second graph: This is a plot of Hardness vs. Displacement Into
Surface for all tests performed on the sample.
Tip Information Table: This table displays the Tip Name.
Calculation Inputs Table: This table displays important inputs
that were used in the calculations.
Test Inputs Table: This table displays important inputs that were
used to determine how the test was performed.
Quick Start Guide 2