Agilent Nano Indenter G200 User's Guide
In the formula window, highlight the placeholder text slope 1 index.
36
In the left pick list, select Index Points.
37
In the right pick list, double-click on CurrentIndex.
38
Immediately following that text, type -5.
39
Following the same procedure, replace the slope 2 index placeholder
40
text with CurrentIndex.
Immediately following that text, type +5.
41
Place the cursor at the end of the current formula and type a forward
42
slash /.
In the left pick list, select Channels.
43
In the right pick list, double-click on Displacement Into Surface.
44
Verify that the formula is identical to the one shown in
45
Figure D-64Correct formula
Click the Check button.
46
The response should be "Formula is okay."
Click OK.
47
This formula creates a new channel that holds the value of the slope
divided by the current displacement. The slope is calculated based on
the five previously acquired data points and the five succeeding data
points (calculated post-test, not in real time).
NanoSuite Explorer D
Figure
D-64.
D-52