Agilent Nano Indenter G200 User's Guide
Locate your sample. Avoid touching the sample with your fingers,
4
using tweezers if necessary.
Place the sample on the sample disk, as shown in
5
page 2-13.
Figure 2-13Placing the sample on the sample disk
Using wooden contacts (such as toothpicks), press down on the
6
sample so that it seats in the crystalbond, as shown in
Figure 2-14Seating the sample into the crystalbond
Use pliers to remove the sample disk from the heating element and
7
allow the disk to cool to room temperature.
Turn the heating element off.
8
For samples that are sensitive to heat, follow these steps:
Locate a sample disk and place it on a heating element.
1
Turn the heating element on medium-low heat and wait for the disk
2
to heat for about 5 minutes.