Frame Stiffness and Area Function Calibration
General Information
Agilent Nano Indenter G200 User's Guide
The first step in calibrating the frame stiffness and area function is to
acquire data on fused silica over a range of displacements.
There are three methods from which the user may determine machine
compliance. These methods are:
•
Frame Stiffness and Area Function Calibration: Express Test
rapid indentation.
•
Frame Stiffness Calibration: ISO 14577 Method
indentation.
•
Frame Stiffness Calibration: CSM Method
indentation.
See
Frame Stiffness and Indenter Area Function Theory
"Theory" for more theoretical information.
Frame stiffness and area function information are stored in the same
location. To access the information, choose Tip > Configure Tip from
the main menu as shown in
Figure 6-45Select Configure Tip from the Tip menu
Tip definitions are stored by user-defined names. The recommended
convention to use is Serial No. Date Method (ie TB12938 042812
Express Test) as shown in
Commonly Used Procedures 6
using dynamic
Figure
6-45.
Figure 6-46
on page 6-32.
using
using quasi-static
in
Chapter
7,
6-31