Frame Stiffness Calibration: ISO 14577 Method

Agilent Nano Indenter G200 User's Guide
Make sure that the fused silica standard is in the center sample-disk
1
hole in the sample tray.
In NanoSuite, select Open Method... from the Method menu.
2
From the Filename drop-down box in the Open Method dialog box,
3
select ISO Standards > XP ISO 14577 Standard Test Method as
shown in
Figure
6-72.
Figure 6-72Selecting the ISO method
Click OK.
4
Select Batch Mode from the Mode menu.
5
Right-click in the right pane in NanoSuite and select Load Batch.
6
In the Retrieve Batch File dialog box, open C:\Program Files\
7
Agilent NanoSuite\Methods\ISO Standards\Machine
Compliance Batch.bch.
Click Open.
8
Click Run Batch (see
9
Commonly Used Procedures 6
Figure 6-73
on page 6-52) to initiate testing.