Testing the DCM II Performance

Agilent Nano Indenter G200 User's Guide
Select the name of the calibrated tip from the Tip Selection
2
drop-down menu in the Select Tip dialog box shown in
on page B-15.
Figure B-22Select name of calibrated indenter tip
If a resulting Select NanoTip confirmation box appears, click OK to
3
dismiss it.
After installing a new DCM tip, the machine compliance must be
determined. Follow the procedure in
Procedures.
Once the Frame Stiffness and Area Function calibrations are completed,
continue to
Testing the DCM II
Select Open Method... from the Method menu.
1
In the Open Method dialog box, select DCM > DCM Standard
2
Hardness, Modulus, and Tip Cal, as shown in
page B-16.
Dynamic Contact Module II B