Failure Of The Integrated Patient-Gas Measurement Module (Pgm) - Dräger Atlan A100 Instructions For Use Manual

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Troubleshooting
Cause
The O
ing.
12.7
Failure of the integrated patient-gas measurement
module (PGM)
WARNING
Risk due to gas measurement failure
If the gas measurement fails, the patient can no longer be adequately monitored.
► Check the sample line and the water trap for damage or blockages. Pay
attention to the replacement intervals.
► Replace the patient-gas measurement module.
► Ensure corresponding substitute monitoring.
● Arrange for appropriate substitute monitoring conforming to ISO 80601-2-55.
12.7.1
Replacing the patient-gas measurement module (PGM)
Contact service personnel to replace the PGM.
Prerequisite:
– There is no patient connected.
– The device is switched off.
– The device is disconnected from the power supply. To do this, disconnect the
power plug.
– The water trap has been removed.
WARNING
Incorrect gas measurement
Faulty or non-functional PGMs can cause incorrect gas measurements. As a result,
the patient could be put at risk.
► Have the PGM replaced by service personnel.
1. At the left-hand side of the device, take hold of the cover (1) at the top and
bottom and pull it off directly to the front.
238
sensor cell is faulty or is expir-
2
Instructions for use
Remedy
Replace the O
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
sensor cell.