Assembly and preparation
Silicone residues or aerosol residues in the water trap can get into the measuring
cuvette. As a result, the measurement may be compromised or a fire could start,
putting the user and patient at risk.
► Do not spray the O-rings of the water trap holder with silicone spray.
4.1.4
Electrical safety
Ambient conditions
If the device is operated at ambient temperatures above 35 °C (95 °F), the battery
cannot be charged properly. The power supply out of the battery may be limited. As
a result, the patient could be put at risk.
► Do not expose the device to temperatures above 35 °C (95 °F) on a permanent
basis.
Mains power supply
If the device is connected to a power socket with incorrect mains voltage or a power
socket without a protective ground, an electric shock may occur. Connecting the
device to power socket strips can lead to an increased leakage current. This may
result in an electric shock and the user and patient may be put at risk.
► Connect the device only to power sockets with correct mains voltage and a
protective ground.
► Do not connect the device to power socket strips.
Battery supply
If the battery is not sufficiently charged, it may not be possible to maintain operation
for long enough if the mains power supply fails. As a result, the patient could be put
at risk.
► Before first operation or after storage, charge the battery for at least 8 hours.
► Check the functional integrity of the battery by performing regular inspections.
Auxiliary power sockets
The connection of devices to auxiliary power sockets can lead to an increased
leakage current. If the protective ground conductor of one of these devices fails, the
leakage current may rise above the permissible values. This may result in an
electric shock or device failure and the patient and user may be put at risk.
► Do not connect additional power socket strips to the auxiliary power sockets on
the device, but rather to separate power sockets on the wall.
► Have service personnel check the leakage current, especially if additional
devices will be connected.
► If the permissible leakage current is exceeded, use separate power sockets on
the wall instead of the auxiliary power sockets on the device.
► Do not connect high-frequency surgery equipment to the auxiliary power
sockets.
► When making a connection, follow the manufacturer's instructions for all
connected devices.
62
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n