Dräger Atlan A100 Instructions For Use Manual page 280

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Technical data
Compliance (without breathing circuit)
In the Man / Spon mode (including
disposable CO
absorber)
2
In mechanical ventilation modes
Total leakage of the breathing system
Filling volume of the CO
Reusable CO
absorber
2
Disposable CO
absorber CLIC absorber
2
800+
Disposable CO
absorber CLIC absorber
2
Free
Flexible breathing bag arm
Volume
Compliance
Rigid breathing bag arm
Volume
Compliance
APL valve
Setting range
Accuracy (at a flow of 20 ±1 L/min):
Pressure drop at 30 L/min (ATPD), fully
opened
280
2.7 ±0.2 mL/hPa (mL/cmH
corresponds to 81 ±6 mL at 30 hPa (cmH
for the "Neonates" patient category
2.9 ±0.2 mL/hPa (mL/cmH
corresponds to 87 ±6 mL at 30 hPa (cmH
for the "Pediatric patients" patient category
3.0 ±0.2 mL/hPa (mL/cmH
corresponds to 90 ±6 mL at 30 hPa (cmH
for the "Adults" patient category
1.0 ±0.2 mL/hPa (mL/cmH
corresponds to 30 ±6 mL at 30 hPa (cmH
for the "Neonates" patient category
1.2 ±0.2 mL/hPa (mL/cmH
corresponds to 36 ±6 mL at 30 hPa (cmH
for the "Pediatric patients" patient category
1.3 ±0.2 mL/hPa (mL/cmH
corresponds to 39 ±6 mL at 30 hPa (cmH
for the "Adults" patient category
<150 mL/min at 30 hPa (cmH
BTPS conditions
absorber
2
1500 mL ±50 mL
1300 mL ±50 mL
1200 mL ±50 mL
0.11 L ±50 mL
0.11 ±0.05 mL/hPa (mL/cmH
corresponds to 3.3 ±1.5 mL at 30 hPa (cmH
0.13 L ±50 mL
0.13 ±0.05 mL/hPa (mL/cmH
corresponds to 3.9 ±1.5 mL at 30 hPa (cmH
Open, 5 to 70 hPa (cmH
±20 % of the set value or ±3 hPa (the greater value
applies), but not more than +10 hPa (cmH
Dry: 2.1 ±0.3 hPa (cmH
Wet: 2.2 ±0.3 hPa (cmH
Instructions for use
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
O)
2
O)
2
O)
2
O)
2
O)
2
O)
2
O) standardized to
2
O)
2
O)
2
O)
2
O)
2
O)
2