Dräger Atlan A100 Instructions For Use Manual page 293

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Gas supply connection
Gas cylinders (dimensions)
Diameter
Maximum height
Pressure reducer (permanently fitted)
Version
Permissible inlet pressure (P
Air, O
2
N
O
2
Opening pressure of the relief valve
Pressure reducer (third-party manufac-
turer)
Version
Noise emissions from device
Measuring method
Equivalent continuous sound pressure level
Leq(A) during ventilation with typical settings
Sound pressure L(A) of alarm tones at the
user's operating location
Measuring method
Acoustic alarm signal
Alarm volume (all priorities)
Secondary acoustic alarm signal and
mains power supply failure alarm
Dimensions of the device, including
breathing system and ventilator
Information
Compact version with plug-in connector for
1 vaporizer (may deviate with accessory
equipment)
Width
Height
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Depending on configuration: NIST, DISS (CGAV-5/B or
CGAV-5/N), French standard (NFS90-116)
100 to 140 mm (3.94 to 5.51 in) for devices equipped
with upright gas cylinders
100 to 110 mm (3.94 to 4.33 in) for devices equipped
with hanger yoke system for gas cylinders with pin-
index connections
880 mm (34.64 in) for devices equipped with upright
gas cylinders
757 mm (29.80 in) for devices equipped with hanger
yoke system for gas cylinders with pin-index connec-
tion
In accordance with ISO 80601-2-13
)
V
Up to 200 kPa x 100 [bar] (2900 psi, 20 MPa)
Up to 60 kPa x 100 [bar] (870 psi, 6 MPa)
7.0 to 9.5 kPa x 100 [bar] (101 to 138 psi, 0.70 to
0.95 MPa)
In accordance with DIN EN ISO 10524-1
Free field measurements complying with ISO 3744
≤42 dB(A)
In accordance with IEC 60601-1-8
Adjustable from ≥42 dB(A) to <80 dB(A)
≥50 dB(A) and ≤80 dB(A)
A tolerance of ±10 mm (0.39 in) applies to all specified
dimensions.
745 mm (29.3 in)
1403 mm (55.2 in)
Technical data
293