11.5.6.4
Vertical tab "System test"
Headline/
Parameter
General
Always use walk-
through mode
Verify O
delivery
2
and circuit
(requires sample
line)
Test gas supply
Central O
supply
2
Central Air supply
Central N
O sup-
2
ply
O
cylinder
2
Air cylinder
N
O cylinder
2
11.5.6.5
Vertical tab "Logbook"
Headline/
Parameter
A logbook entry with measured values is created
Every
1 min; 2 min; 5 min; 10 min; 15 min
For all high-
priority alarms
For all medium-
priority alarms
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
On; Off
On; Off
On; Off
On; Off
On; Off
On; Off
Setting range
On; Off
On; Off
Description
Sets the walk-through mode function.
When On is configured, system tests
and leakage tests will always be exe-
cuted in walk-through mode.
Specifies whether the automatic tests
will check whether the O
ally delivers oxygen and whether the
breathing circuit is correctly connected.
Prerequisite: The sample line is con-
nected to the Y-piece or the filter on the
Y-piece during the tests.
If On is configured, the device automat-
ically checks the O
supply and the cor-
2
rect connection of the breathing circuit.
If Off is configured, the O
checked. In order to verify the correct
connection of the breathing circuit,
additional manual test steps must be
completed in the checklist.
Specifies which gas supplies are tested
during the automatic system test.
If Auto On is configured, only the gas
supplies are checked whose actual
availability is guaranteed after auto-
matic start-up.
At least one O
source must be tested
2
during the system test.
Description
Enables the creation of additional log-
book entries with measured values.
Generates periodic entries.
Generates entries in the event of an
alarm.
Configuration
supply actu-
2
supply is not
2
219

Advertisement