8. Secure the gas cylinders (4) with hook-and-loop straps (not shown here).
Replacing the gas cylinder
1. Remove the old sealing washer (1).
2. Insert a new sealing washer (1) on the cylinder holder (2).
3. Continue with fitting the gas cylinders, see "Fitting the gas cylinders".
If required, the gas cylinder valve (3) can be opened with the supplied wrench (4).
4.5.2.3
Handling O
WARNING
Risk of explosion
When pressurized, O
► Do not oil or grease the gas cylinder valve or the pressure reducer on the O
cylinder. Do not touch with oily or greasy fingers.
● The gas cylinder valves must only be opened and closed slowly. Do not use any
tools on equipment with screw connections.
Have service personnel replace any leaky or stiff gas cylinder valves.
4.5.3
Fitting the vaporizers
The device can be operated with vaporizers which have a Dräger Auto Exclusion,
Interlock or Selectatec plug-in adapter. Dräger recommends using only vaporizers
that are listed in the list of accessories and have been tested.
The vaporizers used must conform to the ISO 8835-4 or ISO 80601-2-13 standard.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
gas cylinders
2
is self-igniting in combination with oil or grease.