Dräger Atlan A100 Instructions For Use Manual page 123

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Complete leakage test
Prerequisite: The Verify O
deactivated in the system setup (see "Vertical tab "System test"", page 219).
Testing the inner hose:
1. Connect the breathing circuit as follows:
a. Connect the hose (1) to the inspiratory port (2).
b. Connect the coaxial test adapter (3) to the expiratory port (4).
c. Remove the elbow (5) from the hose (6).
d. Connect the hose (6) to the coaxial test adapter (3).
2. Touch the Tests... button.
3. Touch the Leakage test button.
The determined leakage value is displayed as Leakage (mech. vent.). Note down
the value as the leakage of the inspiratory hose.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
2
2
1
delivery and circuit (requires sample line) setting is
Tests
4
3
5
6
123