2. Touch the Stop button (2) to stop.
3. To reset the volumeter and time display to zero, touch the Reset button (2).
The volumeter stops automatically after 60 seconds. The measured values are
displayed for 4 minutes and then deleted.
8.6.6
Low-flow wizard
Prerequisite: The device has the "advanced gas monitoring" software option.
The low-flow wizard displays bar graphs for the required fresh-gas flow and the
current total flow. Both bar graphs are to the same scale.
Low-flow wizard
Required FG flow
Total flow
An evaluation of the total flow is displayed below the bar graph:
Evaluation
Too high
Efficient
Too low
Refill bag
An insufficient fresh-gas flow may trigger the Fresh gas low or leakage or
Ambient air inlet activated alarms, for example.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Too high
Color
Yellow
Green
Red
Red
Meaning
The fresh-gas flow is possibly too high. If
the fresh-gas flow can be reduced, both
fresh gas and anesthetic agent will be
saved.
No action is necessary.
The fresh-gas flow is too low.
● Check the fresh-gas flow.
● Check the position of the breathing
bag.
The fresh-gas flow is too low.
● Check the filling of the breathing bag. If
necessary, fill up the breathing bag,
e.g., with the O
flush.
2
Operation
167