Table of Contents

Advertisement

4. Press the O
pressure exceeds 15 hPa (cmH
5. Pull off the Y-piece and plug it on again. The pressure drops.
7.5.4
Valves
7.5.4.1
Checking the O
1. Press the O
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
+ key. The inflow of gas is audible. Keep the key pressed until the
2
O).
2
flush
2
+ key until the pressure exceeds 15 hPa (cmH
2
Tests
O).
2
131