Suction Line Filter; Gas Supply Module (Optional); Apgar Timer; Weigh Scale - Dräger Resuscitaire WBR82 Instructions For Use Manual

Radiant warmer product
Hide thumbs Also See for Resuscitaire WBR82:
Table of Contents

Advertisement

PROPRIETARY AND CONFIDENTIAL DRAFT 5 Feb 04

Suction Line Filter

The Suction Line Filter is a hydrophobic bacterial filter with
extension tubing. It is connected in line with the supply
connection to the suction bottle. The extension tubing
connects to the suction bottle outlet port.

Gas Supply Module (Optional)

The gas supply module includes an On/Off switch that
controls the pipeline and cylinder gas supply to the
resuscitation module. The basic resuscitation module
includes an oxygen cylinder pressure gauge.
NOTE:
When both wall and cylinder gas sources are turned on, the
system may not always pull gas from the wall source. It
pulls from the source with the highest pressure. Check cylinder gauges to determine gas source.
On units equipped with the blender module, oxygen and air
pressure gauges are provided.
NOTE:
The Blender Module option changes the Resuscitation
Module to add the air pipeline and the cylinder connections.
It also adds an air gauge to the Gas Supply Module.

APGAR Timer

When the APGAR Timer is activated, the APGAR Timer display shows elapsed minutes and seconds up
to 59:59, and a notification tone sounds at the 1, 5, and 10-min intervals.

Weigh Scale

The scale consists of a Display Module and Weighing Platform. A microcomputer is used to determine
the weight of infants in the Resuscitaire Birthing Room Warmer and the Resuscitaire Radiant Warmer.
The infant weight remains on the digital display for several minutes. Refer to the User Manual that
accompanies the scale for additional information.
2 - 10

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Resuscitaire rw82Resuscitaire wmrw82

Table of Contents