Dräger Atlan A100 Instructions For Use Manual page 265

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Priority
Alarm
!!! 100
System failure
!!! 0
Therapy settings not
applied
!! 100
Third agent detected
!! 90
Tidal volume not achieved The delivered inspiratory
!
0
USB write error
!!! 120
Ventilator failure
!
160
Volume inaccurate (ambi-
ent pressure implausible)
!
100
Water trap disconnected? Infinity ID water trap is not
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Cause
The gas delivery and the
controlled ventilation have
failed due to an internal fault.
The last changes to the ther-
apy settings were not
applied.
A mixture of more than 2
anesthetic agents has been
detected.
There is electromagnetic
interference.
tidal volume is lower than the
set value.
The USB flash drive is full,
faulty, write-protected, or not
compatible. The USB flash
drive is not correctly con-
nected. The USB flash drive
is not correctly formatted.
The piston ventilator or at
least one pressure sensor is
faulty.
The ambient pressure mea-
surement is implausible. Vol-
ume monitoring and volume
delivery may be inaccurate.
correctly connected.
Troubleshooting
Remedy
Proceed as follows: -Activate
the backup manual mode,
then ventilate the patient with
the breathing bag. Or -Dis-
connect the patient, then use
a manual resuscitator.
Call Dräger.
Change to Man/Spon mode,
then switch back to the
desired ventilation mode.
Use "ALARM RESET" to
reset alarm.
If the problem persists, call
Dräger.
Flush the system if neces-
sary.
Check fresh-gas settings.
Wait for transition phase to
end.
Check for electromagnetic
radiation in the vicinity.
Check ventilation settings.
Check Pmax setting.
Check patient compliance.
Check if the patient is breath-
ing spontaneously.
Check the USB flash drive.
Use "ALARM RESET" to
reset alarm.
Start manual ventilation!
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
Use "ALARM RESET" to
downgrade alarm priority.
Check the setting for "Height
above sea level" in the sys-
tem setup.
Perform the system test.
If the problem persists, call
Dräger.
Check water trap.
265