Dräger Atlan A100 Instructions For Use Manual page 283

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Volume
Information
Expiratory tidal volume VT
Range
Lowest detectable VT
Accuracy
Resolution of displayed value
Minute volume
Total MV
Mandatory MVmand
Spontaneous MVspon
Range
Accuracy
Resolution of displayed value
T
10...90
Minute volume MV
Range
Accuracy
Resolution of displayed value
T
10...90
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
The measured volume values displayed already take
account of the compliance of the breathing circuit.
For devices with the integrated patient-gas measure-
ment module (PGM), the determined gas composition
and the sample gas flow of the gas measurement are
also taken into consideration as soon as CO
tory phases are detected.
For devices without integrated PGM, the specified
accuracies apply if the flow correction is correctly set
(see the "Flow correction" chapter).
The accuracies specified apply under test conditions in
accordance with ISO 80601-2-13 (ventilation with dry
oxygen, corrected to BTPS conditions).
0 to 2500 mL
≤3 mL for the "Pediatric patients" and "Neonates"
patient categories
≤20 mL for the "Adults" patient category
±4 mL in the range below 20 mL
±10 % of the measured value or ±4 mL (the greater
value applies) in the range from 20 to 150 mL
±10 % of the measured value or ±15 mL (the greater
value applies) in the range above 150 mL
1 mL
0.00 to 40.0 L/min
±10 % of the measured value or ±0.15 L/min (the
greater value applies)
0.01 L/min (for MV <1 L/min) or 0.1 L/min (for MV
≥1 L/min)
<45 s (for RR ≥6 /min)
<105 s (for RR <6 /min)
0.00 to 40.0 L/min
±10 % of the measured value or ±0.15 L/min (the
greater value applies)
0.01 L/min (for MV <1 L/min) or 0.1 L/min (for MV
≥1 L/min)
<45 s (for RR ≥6 /min)
<105 s (for RR <6 /min)
Technical data
respira-
2
283