Dräger Atlan A100 Instructions For Use Manual page 216

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Configuration
Headline/
Parameter
Auto
Air
N
O
2
Enable pause
mode
11.5.5
System setup > Licenses/Options
11.5.5.1
Vertical tab "Licenses/Options"
Headline/
Parameter
Licenses for software options
Gas supply
Disable N
O
2
216
Setting range
On; Off
Setting range
On; Off
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Description
Corrects the measured flow values
automatically when an anesthetic gas
monitor is connected.
Corrects the measured flow values
when Air is used.
Corrects the measured flow values
when N
O is used.
2
Defines whether the Pause mode is
available.
On: Pause mode is activated.
Off: Pause is deactivated.
For devices with integrated patient-gas
measurement module, Monitoring
mode is available as an alternative.
Description
– Overview of available and active
software options.
– Activating software options.
Observe the information in the following
section: "Activating software
options", page 224
A device with electronic gas mixer and
connectors for nitrous oxide can be
configured so that nitrous oxide is no
longer displayed and cannot be
selected as a carrier gas.
Prerequisite:
– The current carrier gas is Air.
– Nitrous oxide is not connected or
available.
Restart the device if necessary to
implement the settings.
On: Nitrous oxide cannot be delivered.
Off: Nitrous oxide can be delivered.