Troubleshooting
Priority
Alarm
!
170
Power failure
!! 100
Power supply failure
!! 100
Pressure sensor failure
!
255
Pressure-relief valve
opened
!!! 200
Rotary knob stuck
!
170
Sample line disconnected? The sample line or the water
!
170
Sample line occluded
!
75
Second agent detected
!
60
Service date reached
!!! 200
Set O
switch to "Add. O
2
!! 0
Speaker failure
264
Cause
Mains power supply is not
available. The device has
switched to battery opera-
tion.
Internal fault in the power
supply.
There is condensate in the
breathing hoses.
Sensor calibration failed.
The internal pressure-relief
valve has been activated due
to overpressure.
Key is stuck or was pressed
for more than 10 seconds.
trap has been disconnected.
Sample line is occluded.
A second anesthetic agent
has been detected.
Maintenance is due.
" The internal gas delivery sys-
2
tem has failed and the O
switch is still set to "Aux. O
The loudspeaker is faulty.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Remedy
Restore mains power supply.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
Check the breathing hoses.
Perform the system test.
If the problem persists, call
Dräger.
Check APL valve and fresh-
gas settings.
The therapy will continue with
current settings.
Press and turn the rotary
knob repeatedly.
If necessary, proceed as fol-
lows: -Activate the backup
manual mode, then ventilate
the patient with the breathing
bag. Or -Disconnect the
patient, then use a manual
resuscitator.
Check sample line and water
trap.
Check sample line, water
trap, and patient-side filter.
Flush the system if neces-
sary.
Check fresh-gas settings.
Wait for transition phase to
end.
Use "ALARM RESET" to
reset alarm.
Call Dräger.
Set the O
switch to "Add.
2
O
". Set the O
flowmeter to
2
2
2
".
the desired flow.
2
Call Dräger.
Use "ALARM RESET" to
downgrade alarm priority.