Dräger Atlan A100 Instructions For Use Manual page 250

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Troubleshooting
Priority
Alarm
!! 180
Battery charge low
!!! 30
Battery charge very low
!! 170
Battery failure
!
100
Battery temperature high
!
100
Breathing circuit expired
250
Cause
The battery charge is low
and the mains power supply
is not available.
The battery charge is critical
and the mains power supply
is not available. The device
may shut down in the next
minutes.
The battery is faulty. If the
mains power supply fails, the
device will switch off immedi-
ately.
The battery temperature is
high. Charging of the battery
has been suspended to pro-
tect it from damage.
Accessory has been used
too long.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Remedy
Make sure that the mains
power supply is correctly
connected.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Use "ALARM RESET" to
downgrade alarm priority.
Restore mains power supply.
The breathing system
warmer has been switched
off. Check the breathing cir-
cuit for condensate. Increase
the fresh-gas flow if neces-
sary.
Once the battery has been
depleted, ventilate the patient
manually.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
Ensure that the system is
connected to the mains
power supply.
Check the ambient tempera-
ture.
Use "ALARM RESET" to
reset alarm.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
reset alarm.