Dräger Atlan A100 Instructions For Use Manual page 275

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Resolution of the value displayed in the
status display
Total flow tube
Standardization
Range
Accuracy
Resolution of displayed value
15.6
Fresh-gas delivery with a mechanically controlled gas
mixer and flow tubes
Standardization
O
concentration
2
Setting range for fresh-gas flow
Measurement of the fresh-gas flow for O
and N
O
2
Range
Accuracy
Resolution of displayed value
Measurement of the fresh-gas flow for Air
Range
Accuracy
Resolution of displayed value
15.7
Ventilator
Design
Time-based settings
Respiratory rate RR
Setting range
Increment value
Accuracy
Minimum respiratory rate RRmin in PSV mode
Setting range
Increment value
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
0.1 L/min
All data are standardized to STPD conditions.
Off to 10 L/min
±10 % of the set value at 100 % O
>1 L/min
0.5 L/min (in the range from 1 to 10 L/min)
All data are standardized to STPD conditions.
21 to 100 Vol%
Off to at least 10.0 L/min (O
2
0.02 to 10.0 L/min
±10 % of the delivered flow (from 0.1 to 10.0 L/min)
±50 % of the delivered flow (below 0.1 L/min)
0.05 L/min (from 0.05 to 1.00 L/min)
0.50 L/min (from 1.00 to 10.0 L/min)
0.2 to 10.0 L/min
±10 % of the delivered flow (from 0.6 to 10.0 L/min)
±50 % of the delivered flow (below 0.6 L/min)
0.2 L/min (from 0.2 to 1.0 L/min)
0.5 L/min (from 1.0 to 6.0 L/min)
2.0 L/min (from 6.0 to 10.0 L/min)
Electronically driven ventilator, fresh-gas decoupled
3 to 100 /min
1 /min
±10 % of the set value or ±1 /min (the greater value
applies)
Off, 3 to 60 /min
1 /min
Technical data
and for flows
2
, Air, and N
O)
2
2
275