11.5.6
System setup > System
11.5.6.1
Vertical tab "General"
Headline/
Parameter
Language
Height above sea
level [m]; [ft]
Time source
MEDIBUS 1; MEDIBUS 2; NTP server;
Date and time
Automatic switch
to daylight sav-
ings time
OR working hours
Device name
Device name (up to 16 alphanumeric
User password
Reset all pages to
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
List of available languages
English (United States)
[m]: 0 up to 4000
[ft]: 0 up to 13120
None
day; month; year
hour; minute
On; Off
Hour : Minute to Hour : Minute
6:30 to 18:30
characters)
A3XX
Change password
Factory defaults
Description
Selects the language.
A flag symbol identifies the tabs that
lead to the page with the language set-
tings.
Sets the height above sea level.
Selects the source for the time synchro-
nization.
Prerequisite: The connected device
supports this function.
Sets the date and time. The change is
applied on leaving the "General" verti-
cal tab.
Activates or deactivates the automatic
changeover to daylight saving time.
Sets the working hours of the operating
room.
During this time, the gas measurement
is kept in a pre-warmed and calibrated
state so measured values are available
after only a short waiting period. How-
ever, this decreases the life span of the
patient-gas measurement module.
Changes the device name in order,
e.g., to enter the operating location.
Changes the user password.
Resets all the settings on all pages in
the System setup dialog to the factory
defaults.